PLATE and MESH PREPARATION
1)
Thermo drapes for the Inion Thermo water bath are supplied
sterile. The water bath takes 20-30 minutes to heat the sterile
water to the required temperature of 55C. Inion CPS plates
and meshes take approximately one minute to be fully activated
in the water bath. Plate and mesh adaptation is carried out
after activation.

2)
After activation the plates retain a degree of malleability
for several hours, during which time modifications can still
be made. The plates can be reactivated at any time during
the procedure by re-immersion in the water bath for an additional
15-60 seconds.

Thermo drapes for the Inion Thermo water bath are
supplied sterile.
The water bath takes 20-30 minutes to heat the sterile water
to the required temperature of 55°C.
Inion CPS plates and meshes take approximately one minute
to be fully activated in the water bath.
Plate and mesh adaptation is carried out after activation.
The plates can be reactivated at any time during the procedure
by re-immersion in the water bath for an additional 15-60
seconds.
|